Electron microprobe



  An electron microprobe (EMP), also known as an electron probe microanalyser (EPMA) is an analytical tool used to non-destructively determine the chemical composition of small volumes of solid materials. It works similarly to an electron microscope, in which the sample is bombarded with an electron beam. In EPMA, however, the resulting parts per million (ppm).

History

Development of the electron microprobe was preceded by that of the closely related analytical technique of Georg von Hevesy in 1923 and applied by other workers in the following few years.

In 1944, MIT built an electron microprobe, combining an electron microscope and an energy-loss spectrometer. Electron energy-loss spectrometry is very good for light element analysis and they obtained spectra of C-Kα, N-Kα and O-Kα radiation. In 1947, Hiller patented the idea of using an electron beam to produce analytical X-rays, but never constructed a working model. His design proposed using Bragg diffraction from a flat crystal to select specific X-ray wavelengths and a photographic plate as a detector.

In 1948-1950, quartz crystal between the sample and the detector to permit wavelength discrimination. He also added an optical microscope to view the point of beam impact. The resulting microprobe was described in Castaing's 1951 Ph.D. thesis, in which he laid the foundations of the theory and application of quantitative analysis by electron microprobe, establishing the theoretical framework for the matrix corrections of absorption and fluorescence effects. Castaing (1921-1999) is considered the "father" of electron microprobe analysis.

Cameca (France) produced the first commercial microprobe, the “MS85,” in 1956. It was soon followed by many microprobes from other companies; however, all companies except Cameca and JEOL, are now out of business. In addition, many researchers build electron microprobes in their labs. Significant subsequent improvements and modifications to microprobes included scanning the electron beam to make X-ray maps (1960), the addition of solid state EDS detectors (1968) and the development of synthetic multilayer diffracting crystals for analysis of light elements (1984).

Process

Low-energy electrons are produced from a tungsten filament cathode and accelerated by a positively biased Auger electron production.

The characteristic X-rays are used for chemical analysis. Specific X-ray wavelengths are selected and counted, either by energy dispersive X-ray spectroscopy (EDS). WDS utilizes Bragg diffraction from crystals to select X-ray wavelengths of interest and direct them to gas-flow or sealed proportional detectors. In contrast, EDS uses a solid state semiconductor detector to accumulate X-rays of all wavelengths produced from the sample.

Chemical composition is determined by comparing the intensities of characteristic X-rays from the sample material with intensities from known composition (standards). Count from the sample must be corrected for matrix effects (absorption and secondary fluorescence) to yield chemical compositions. The resulting chemical information is gathered in textural context. Variations in chemical composition within a material (zoning), such as a mineral grain or metal, can be readily determined.

See also

References

  1. ^ Wittry, David B. (1958). "Electron Probe Microanalyzer", US Patent No 2916621, Washington, DC: U.S. Patent and Trademark Office.
 
This article is licensed under the GNU Free Documentation License. It uses material from the Wikipedia article "Electron_microprobe". A list of authors is available in Wikipedia.